v Easy to replace Filter in front panel
v Ultra HEPA Filter 99.999% of 0.3µm Particle Remove
v This contaminated air is drawn by the blower through the back plenum of the cabinet, where approximately 70% of the air is re-circulated through the supply HEPA filter and back over the work area.


This contaminated air is drawn by the blower through the back plenum of the cabinet,
where approximately 70% of the air is recalculated through the supply HEPA filter
and back over the work area.
Ultra HEPA Filter 99.999% of 0.3 Particle Remove
|
Model
|
LCB-0103B-A2
|
LCB-0123B-A2
|
LCB-0153B-A2
|
|
Dimensions
|
Inner
|
900 x 600 x 700
|
1200 x 600 x 700
|
1500 x 600 x 700
|
|
(W x D x H)
|
Outer
|
1030 x 810 x 1500
|
1360 x 810 x 1500
|
1670 x 810 x 1500
|
|
Air Flow Velocity
|
|
0.35~0.5 M/Sec
|
|
Air Flow Volume
|
|
980 m²/h
|
1100 m²/h
|
1900 m²/h
|
|
Exhaust Filter
|
|
>99.999% Effiency ULPA Filters >0.3µΜ
|
|
Down Flow Filter
|
|
>99.999% Effiency ULPA Filters >0.3µΜ
|
|
Micro Process
|
|
Air Velocity, Cabinet temperature, UV Lamp hour meter, Filter change date
|
|
Controller With
|
|
Visual & audible Alarm for Low Air flow & Velocity, Flow Malfunction
|
|
LCD Display
|
|
Microprocessor increases fan speed gradually in filters resistance
|
|
Material
|
Inner
|
Stainless Steel AISI 301
|
|
|
Outer
|
Electro- galvanized steel with epoxy power coating
|
|
Door
|
Laminated & Tempered safty front glass window
|
|
IIIumination
|
|
>1100 Lux
|
|
Noise Level
|
|
<55 dBA
|
|
Fans
|
|
Centrifugal high efficiency fan with speed control
|
|
Utility Device
|
|
UV lamp, FL lamp, Air / Gas stop cock, 110/220V Consent
|
|
Electronic Supply
|
|
110 V , 60 Hz or 220 /240V , 50/60 Hz
|
|
Option
|
Movable Stand
|